Mode of Operation
The chamber is cooled with LN2 and heated with electrical resistance elements. Typical cycling times from hot to cold, and from cold to hot, are 45 minutes and 65 minutes, respectively.
Parameters
Temperature Range: -180°C to +200°C
Cryogenic Liquid: LN2 at 50 psig
Heating Capacity: 510 W
Cooling Capacity: 100 W at -150°C
Physical Characteristics
Test Volume: 20" x 12" x 16" (W x H x L)
Viewport Size: 8" x 8"
Power: 120 VAC, 1 Phase, 20 amperes
Controller: Watlow 920 Programmable Controller
Integral Instrumentation
Payload Temperature: 7 thermocouple channels
|